PRIMING
PRIMINGPRIMING
PRIMINGPRIMING
PRIMING
Clear Air from the DUT Tubing
Run the Device Under Test (DUT) to clear air from its tubing
before connecting it to the Module for testing.
STEP 7
9
Connect DUT to Valve
Connect the DUT tubing to the Valve, insuring that there are
no bubbles in the line.
STEP 8
9
Run 10 mL of Distilled Water through DUT
For maximum accuracy, run 10 mL of Distilled Water through
the DUT and Module prior to beginning a test.
STEP 9
9