MAIN SCREENS – There are 7 main screens, 5 Display Screens which have 1, 2, 3, 4 and
5 display zones respectively, a Measurement List Screen which shows available
measurements and the Quick Config Screen which displays the current hardware
configuration. In the Display Screens, each Display Zone can be customized to show the
desired parameter from the following options:
This is the mV measured directly from the RF donut.
This is the converted mA measurement based on the RF
donut mV to mA attenuation ratio.
This is the computed power based on load setting and mA
measured.
This is the maximum mV measured in the buffer.
NOTE: This is shown as absolute value.
This is the difference between maximum mV measured and
min mV measured.
mV Peak /
Peak - to - Peak
This is the ratio of Peak versus
Peak to Peak millivolts.
This is the maximum positive mV measured in the buffer. For
asymmetric waveforms this can determine if the output
polarity is reversed.
This is the ratio of peak to rms of the measured waveform.
ESU-2050P Only
This is the duration that the pulsed waveform is on.
(See Diagram 1)
ESU-2050P Only
This is the duration that the pulsed waveform is off.
(See Diagram 1)
ESU-2050P Only
This is the total cycle time of the pulsed waveform
(i.e. Ton + Toff). (See Diagram 1)
ESU-2050P Only
This is the ratio of the pulse on time (Ton)
versus cycle time (Tcyc). (See Diagram 1)
ESU-2050P Only
This represents the RMS mV over one pulsed cycle.
(See Diagram 1)
ESU-2050P Only
This represents the RMS mA over one pulsed cycle.
(See Diagram 1)
ESU-2050P Only
This represents the RMS Watts over one pulsed cycle.
(See Diagram 1)