CONFigure Subsystem
This group allows the user to setup the display and operational settings for the unit.
KEYWORD PARAMETER FORM COMMENTS
CONFigure
:DISPlay :SxZy S<display_screen_number> Z<zone_number> nn
display_screen_number = 1-7
nn = Parameter for selected Zone
1 = One Parameter
0 = mV RMS
7 = Creast Factor
2 = Two Parameters
1 = mA RMS
8 = Time-Pulse On
3 = Three Parameters
2 = Watts RMS
9 = Time-Pulse Off
4 = Four Parameters
3 = mV Peak
10 = Time-Total Cycle
5 = Five Parameters
4 = mV Pk-Pk
11 = % Duty Cycle
6 = Measurement List Display (Non-editable)
5 = mV Pk / mV Pk-Pk
12 = mV Pulse Cycle
7 = Quick Configuration Screen (Non-editable)
6 = mV Pk+
13 = mA Pulse Cycle
14 = Watts Pulse Cycle
:SCReen < numeric_value >
:AVERaging FAST | SLOW | MEDium
:HOLD ON | OFF
:LOAD
:MODE TABle | ADJustable
:VALue < numeric_value > Table Mode: 1-12 for individual
resistors, 13-22 for resistor
sets 1-10
Adjustable Mode: 0-6500.0
:SETn < numeric_value > n = Resistor set to configure
(1-10)
< numeric_value > = 16 bit
binary value of resistors to
include in set ‘n’ where bit
0 = Load 1, Bit 1 = load 2…
Bit 11 = Load 12
each bit selects whether the
load is included in the set.
Bit = 1 includes the load
Bit = 0 excludes the load.
:NUMber < numeric_value > set of valid numeric values
:Ln < numeric_value > n = Load to configure
< numeric_value > = Actual
value of load ‘n’;
0-6500.0 ohms
INPut:
Atten: 0.1 | 1
RANGe: 100 | 1000 | AUTo
NUMsamples: 1024 | 2048 | 4096 | 8192 | 16384 | 54768
MODE: CONTinous | PULs
Range 1-7
1-5 = # display zones
6=Measurement List Display
7=Quick Config Screen