29
DFA
®
Digital Fast Acquisition Technology is a revolutionary new method of measuring
ESU generator output power. A high-speed analog to digital converter is used to digitize
the high frequency, high power output of the ESU generator. An RF Current Transformer
is used to convert the current signal to a voltage signal, which is read by the analog to
digital converter. By digitizing the signal a more accurate, frequency independent
measurement can be made.
DFA
®
TECHNOLOGY