Patient Simulator Series
Patient Simulator Series
PS-2105
PS-2105
PS-2110
PS-2110
S
S
PECIFICATIONS
PECIFICATIONS
PS-2110
PS-2110
with (optional)
with (optional)
MSP-2100
MSP-2100
& FingerSim
& FingerSim
TM
TM
8
8
ECG Functions
Rate 30,60,80,120,180,240,300 BPM
Accuracy ±1%
Amplitude 0.5, 1.0, 1.5, 2.0 mV
Accuracy ± 2% @ Lead II
High Level 200 times Amplitude
ECG Performance
Sine Wave 0.1,0.5,5,10,40,50,60,100 Hz
Square Wave 0.125, 2.000 Hz
Triangle Wave 2.000 Hz
Rate Accuracy ±1%
Amplitude Accuracy ± 2% @ Lead II
Blood Pressure
Static Pressure
0,20,40,80,100,200,250,300
mmHg
Accuracy ± (2% of Reading + 2 mmHg)
Impedance 300 Ω
Accuracy ± 10%
Excitation Range 2to16VRMS
Excitation Frequency DC to 5 KHz
Sensitivity 5 or 40 µV/V/mmHg
Respiration
Rate
Apnea,15,20,30,40,60,80,100,1
20 BrPM
Accuracy ±1%
Impedance delta 0.1,0.2,0.5,1.0,2.0,3.0 Ω
Accuracy ± 10%
Baseline 500,1000 Ω
Accuracy ±5%
Lead LA or LL
Temperature
Temperature 30,37,40 °C (86.0,98.6,104.0 °F)
Accuracy ± 0.1 °C ( ± 0.1 °F)
Type YSI Series 400 and 700
SpO
2
Simulation (MSP-2100 & Finger SIM
TM
Req’d.)
Rate
30 to 300 BPM
SpO
2
Output 80, 90 & 97%
Flip for product pricing - Page C
Flip for product pricing - Page C
To Compare Patient Simulators,
see chart on Page 6
ECG Functions
Rate 30 to 300 BPM
Amplitude .05 to 5mV
S-T Segment Elevation ± .05 to .8 mV
Artifacts
50 Hz, 60 Hz, Muscle,Baseline
Wander, Respiration
QRS Interval
Adult (80 ms) or Pediatric (40
ms)
Arrhythmia Functions
Arrhythmias 49 different
General Groups
Supraventricular, Premature,
Ventricular, Conduction
Triggering Manual & Automatic
Pacemaker Functions
Waveforms
Atrial, Pacer, Asynchronous,
Non-Capture, Non-Function,
Demand - Occasional, De-
mand - Frequent, AV - Se-
quential
Pulse Height 1to10mV
Pulse Width 0.1 to 2 ms
ECG-Performance Functions
Sine Waves 0.1 to 100 Hz
Square Waves 0.125, 2 Hz
Triangle Waves 5, 2.5 Hz
Pulse Waves 30, 60, 120 BPM; 60 ms Width
Amplitude 0.05 to 5 mV (Lead II)
Automatic Mode Yes
R Wave Rate 30, 60, 80, 120, 200, 250 BPM
R Wave Width 8 to 200 ms
R Wave Amplitude .05 to 5 mV (Lead II)
Training
Timer Manual, 10, 15, 20, 25, 30 s
Randomizer Off, On
Arrhythmias All, Subset
SpO
2
Simulation (MSP-2100 & Finger SIMTM Req’d.)
Rate
30 to 300 BPM
SpO
2
Output 80, 90 & 97%
PS-2105
PS-2105
S
S
PECIFICATIONS
PECIFICATIONS