DPM 225X XX N XX XX
Options
Blank = Not Applicable
DC = DC (Direct Current) Analog Output
HF = High Frequency DC Analog Output
Temperature Sensor Option
N = Not Applicable
Y7 = YSI 700
R1 = 100ȍ RTD
Pressure Range
100 = Max 100 PSI
75 = Max 75 PSI
10 = Max 10 PSI
5 = Max 5 PSI
.3 = Max 0.3 PSI
Model
2251 = Basic Differential Model
2252 = Adds Min/Max and RS-232
DPM 220X XX XX XX XX
Options
Blank = Not Applicable
DC = DC (Direct Current) Analog Output
HF = High Frequency DC Analog Output
Temperature Sensor Option
N = Not Applicable
Y7 = YSI 700
R1 = 100 ȍ RTD
Pressure Range (Sensor 2)
N = Not Applicable
100 = Max 100 PSI
75 = Max 75 PSI
10 = Max 10 PSI
5 = Max 5 PSI
.3 = Max 0.3 PSI
Pressure Range (Sensor 1)
100 = Max 100 PSI
75 = Max 75 PSI
10 = Max 10 PSI
5 = Max 5 PSI
.3 = Max 0.3 PSI
Model
2201 = Basic Model
2202 = Adds Min/Max and RS-232
Digital Pressure Meter Series
Digital Pressure Meter Series
S
S
PECIFICATIONS
PECIFICATIONS
Flip for product pricing - Page N
Flip for product pricing - Page N
M
M
ODEL
ODEL
S
S
UMMARY
UMMARY
BC20 - 21100 BATTERY ELIMINATOR (US Version)
BC20 - 21101 BATTERY ELIMINATOR (Euro Version)
BC20 - 41337 COMMUNICATIONS CABLE (7Pin Mini-Din to DB 9 F)
BC20 - 30106 BC BIOMEDICAL SMALL SOFT SIDED CARRYING CASE
BC20 - 01005 UNIVERSAL PRESSURE ADAPTER KIT
BC20 - 01006 YSI TEMPERATURE CABLE
DPM-2250
DPM-2250
OPTIONAL ACCESSORIES FOR DPM SERIES
32
32
General
Accuracy
Pressure ± .05% FS
Bandwidth ±.5%FS
Display LCD Graphical 128 x 64 Pixels
Setup Memory EEPROM, All Parameters
Memory Retention 10 Years w/o Power
Operating Range 15 to 30 °C
Storage Range -40to60ºC
Construction
Enclosure – ABS Plastic Face –
Lexan, Back Printed
Size 7.09” x 3.94” x 1.56”
Weight ≤1 lb.
Connections
Power – 2.1 mm center Negative
RS232 – DIN Connector Pressure
– Male Luer Temperature – ¼ inch
P
hone
C
AUTION
MEDIA COMPATIBILITY
The DPM-2250 Series Meters are
intended to be used with only non-
corrosive, non-ionic, or otherwise
pure fluids and/or gases that are
compatible with sensor materials
including glass, silicon, ceramic,
epoxy, RTV, gold, aluminum and
nickel
CAUTION
LO PRESSURE PORT LIQUIDS
INCOMPATIBILITY
The Lo port of the DPM-2250 se-
ries meters are not intended for
liquids, only dry gases
Power
BATTERY: 9 V Alkaline LINE: 9 V,
Center Negative
Power Consumption
ON: Less than 12 mA OFF: Less
than 40 µA
Battery Life
CONTINUOUS: 80 hrs OFF: 12
months
Battery Eliminator (Optional)
BC20-21100 (120 V) – US BC20-
21101 (220 V) – Euro 9 V, 200 mA
DC
Analog Output (Optional)
1.0 to 4.0 VDC FS, selectable
+/- .1% FS
DC -- Output dependent on Digital
Filter
HF --Output from 0 to 100 Hz
10 kHz sample rate
Pressure Sensor Ranges
Units 100 75 10
PSIG -13.50 to 100.0 -13.50 to 75.00 -10.00 to 10.00
mmHg -701.0 to 5190 -701.0 to 3893 -519.0 to 519.0
inHg -27.60 to 204.3 -27.60 to 153.2 -20.43 to 20.43
cmH
2
O -951.0 to 7043 -951.0 to 5282 -704.3 to 704.3
inH
2
O -374.0 to 2773 -374.0 to 2080 -277.3 to 277.3
KPa -93.10 to 689.5 -93.10 to 517.1 -68.95 to 68.95
Bar -.9310 to 6.895 -.9310 to 5.171 -.6895 to .6895
mBar -931.0 to 68 95 -931.0 to 51 71 -689.5 to 6 89.5
Units 5 3
PSIG -5.000 to 5.000 -.3000 to .3000
mmHg -259.5 to 259.5 -15.57 to 15.57
inHg -10.22 to 10.22 -.6129 to .6129
cmH
2
O -352.2 to 352.2 -21.13 to 21.13
inH
2
O -138.7 to 138.7 -8.319 to 8.319
KPa -34.48 to 34.48 -2.069 to 2.069
Bar -.3448 to .3448 -.0207 to .0207
mBar -344.8 to 344.8 -20.69 to 20.69
Temperature Ranges
Units 100 Ω RTD YSI 700
D
egrees C
-
20.0 to 100.0
-
20.0 to 100.0
Degrees F -4.0 to 212.0 -4.0 to 212.0