ESU Analyzer Series
ESU Analyzer Series
27
27
Comparison Chart
Comparison Chart
1-888-223-6763 - bcgroupintl.com
1-888-223-6763 - bcgroupintl.com
Models:
ESU-2400 ESU-2300 ESU-2050 ESU-2050P ESU-2000A
Industry-standard
RF current measurement
Yes Yes Yes Yes Yes
Test Loads
1 to 6400 Ohms, 1
ohm increments, in-
ternal load bank
50 to 750 Ohms, 50
ohm increments, in-
ternal load bank
Requires external
precision non-induc-
tive test loads, 1-
9999 Ohms
compatible
Requires external
precision non-induc-
tive test loads, 1-
9999 Ohms
compatible
50, 100, 200, 300,
400, 500 Ohms
RF Current Range <2 ma to 7,000 mA 20 ma to 7,000 mA <2 ma to 7,000 mA <2 ma to 7000 mA 0 - 1000 mA
RF Power (Watts RMS) Range To 999.9 watts To 400 watts To 999.9 watts To 999.9 watts To 500 watts
Bandwidth 10 kHz to 10 MHz 10 kHz to 10 MHz 10 kHz to 10 MHz 10 kHz to 10 MHz 0 - 10 MHz
Display
QVGA Color Touch
Screen
256 x 64 pixel LCD
Graphical Display
with Backlight
128 X 64 Pixels LCD
Graphical with Back-
light
128 X 64 Pixels LCD
Graphical with Back-
light
4.5 inch Analog
Meter
External Test Load
Compatible
Yes, can use internal
+ external or exter-
nal only, or internal
only
Yes, can use internal
+ external or exter-
nal only, or internal
only
Yes, requires exter-
nal test loads
Yes, requires exter-
nal test loads Yes
External Current Transformer
Required
No No Yes Yes No
Continuous Output Waveform
Compatibility
Yes Yes Yes Yes Yes
Pulse Output waveforms
Compatibility
Yes No No Yes Yes
Measurement accuracy
1 % of reading on
RMS Current, 1 %
for reading RMS
Power
2.5 % of reading on
RMS Current, 5 %
for reading RMS
Power
1% reading, > 50
mV,upto1MHz,3%
reading, > 50 mV, 1
to 10 MHz
1% reading, > 50
mV,upto1MHz,3%
reading, > 50 mV, 1
to 10 MHz
+ 5% of Full Scale
Calibration level accuracy Yes No Yes Yes No
Crest Factor Range 1.4 to 500 1.4 to 500 1.4 to 500 1.4 to 500 N/A
Covidien/Valleylab Force Triad
Calibration Procedure
Compatibility
Yes No Yes Yes No
Covidien/Valleylab Ligasure
Generator Calibration
Compatibility
Yes No Yes Yes No
Number of parameters
measured/displayed
15 5 15 15 x 2
Internal REM/ARM/CQM
Testing
Yes, 500 ohm load
adjustable in 1 ohm
increments
Yes, 500 ohm load
adjustable in 1 ohm
increments
No, requires external
CQM Test Box
(CQM-2000)
No, requires external
CQM Test Box
(CQM-2000)
Yes
Automated power-load curves
Yes, "x" to "y" Ohms
in "z" ohm incre-
ments OR by user-
defined load table
values, multiple
power levels @
each load possible
No No No No
Auto-activation of ESU
generator for power load
curve testing
YesNoNoNoNo