OCCLUSION TEST DATALOG OVERVIEW
WORKING WITH THE DATALOGGER
CEnter the Datalogger
CView Occlusion Test Datalog
CView a Specific Record
CGo into Datalog Edit Mode
CScroll to Desired Record Number
CEnter Choice for Record Number
CView or Print Occlusion Test Datalog
DD
DD
D
AA
AA
A
TT
TT
T
ALAL
ALAL
AL
OG--OCCLOG--OCCL
OG--OCCLOG--OCCL
OG--OCCL
USIONUSION
USIONUSION
USION
OCCLUSION TEST DATALOG OVERVIEW
This section covers how to access and choose specific
Infusion Analyzer Occlusion Datalogs. It begins with an
overview and then progresses through step-by-step
instructions for manipulating and using the Datalogs.
The Distal Occlusion tests record the data in the Datalogger
for review and/or generation of a printed report.
The Datalog and report also contain the maximum pressure
point and, if used, the time of the Nurse Call is noted and
three additional pressure readings are taken to insure the
pump was turned off.
No data is recorded for Proximal Occlusion tests.
The following is an outline of the steps that are involved in
working with the Datalogger.