2-25
PRIMING
PRIMING
PRIMINGPRIMING
PRIMINGPRIMING
PRIMING
Begin from a Channel Screen
A Module can be primed when the unit displays the Prime
screen for that Channel. Use to toggle to the desired
prime screen.
STEP 1
9
NOTE:
COnly Distilled Water can be used
with this unit. Do not use tap water,
glucose or any other fluid; this will
cause the interior tubing to become
unclean.
STEP 2
9
CHANNEL 1 IDLE
PRIME WITH 50mL
THEN PRESS START
Attach an External Inlet Valve
To help prevent air from being injected into the Module during
Priming and Testing, attach a Valve to the Fluid In Luer Lock
of the Module. This Valve will be used to minimize the risk of
air being added to the system and will help maintain Prime,
making it unnecessary to prime before every test.
STEP 3
9
Attach a Priming Syringe to the Valve
Attach a syringe filled with 50 mL of Distilled Water to the
Valve.